074_sikkerhed_edwards_perimount

MR-sikkerhed

Hjerteklap-protese / Heart Valve

Carpentier Edwards Perimount

Edwards Lifesciences

Fabrikat/model

Hjerteklap-protese

Heart Valve

Edwards Lifesciences


Denne sikkerhedsvejledning er gældende for følgende modeller...


Carpentier Edwards PERIMOUNT pericardial aortic bioprostheses

Carpentier Edwards PERIMOUNT RSR pericardial aortic bioprostheses

Carpentier Edwards PERIMOUNT Magna pericardial aortic bioprostheses

Carpentier Edwards PERIMOUNT Magna Ease sepericardial aortic bioprostheses         - Carpentier Edwards PERIMOUNT pericardial mitral bioprostheses

Carpentier Edwards PERIMOUNT Plus pericardial mitral bioprostheses

Carpentier Edwards PERIMOUNT Theon pericardial mitral bioprostheses

Carpentier Edwards PERIMOUNT Magne pericardial mitral bioprostheses

Carpentier Edwards PERIMOUNT Magne Mitral Ease pericardial bioprostheses


Statisk feltstyrke

Maksimalt 3,0 Tesla

Scannerarkitektur

Ingen restriktioner

Spatial gradient

Maksimalt 30 T/m (3000 Gauss/cm)

Slew rate (Gr)

Ingen restriktioner

SAR

WB-SAR: 2,0 W/kg ved 15 minutters scanning

Normal Operating Mode

First Mode er ikke tilladt

B1+rms

Ingen restriktioner

Coil type

Ingen restriktioner

Specielle forhold

Patienten kan scannes umiddelbart efter implantation

Temperaturstigning

Maksimalt 2,3 grader ved Normal Operationg Mode i 15 min


Artefaktdannelse

28 - 36 mm fra implantatet

Kilde

Produktinformation fra Edwards Lifesciences

Gyldighed

Denne procedure kan anvendes som inspirationskilde til udvikling af egne sikkerhedsforskrifter

Information

Procedure 074. Revision: 1.0

Udarbejdet af: Bo Haugaard Jørgensen

Review: Overlæge Peter Gideon

Dato: 01.08.2020

Opdateres: Løbende, senest 01.01.2022

Bo Haugaard Jørgensen