Fabrikat/model | Hjerteklap-protese Heart Valve Edwards Lifesciences Denne sikkerhedsvejledning er gældende for følgende modeller... Carpentier Edwards PERIMOUNT pericardial aortic bioprostheses Carpentier Edwards PERIMOUNT RSR pericardial aortic bioprostheses Carpentier Edwards PERIMOUNT Magna pericardial aortic bioprostheses Carpentier Edwards PERIMOUNT Magna Ease sepericardial aortic bioprostheses - Carpentier Edwards PERIMOUNT pericardial mitral bioprostheses Carpentier Edwards PERIMOUNT Plus pericardial mitral bioprostheses Carpentier Edwards PERIMOUNT Theon pericardial mitral bioprostheses Carpentier Edwards PERIMOUNT Magne pericardial mitral bioprostheses Carpentier Edwards PERIMOUNT Magne Mitral Ease pericardial bioprostheses |
Statisk feltstyrke | Maksimalt 3,0 Tesla |
Scannerarkitektur | Ingen restriktioner |
Spatial gradient | Maksimalt 30 T/m (3000 Gauss/cm) |
Slew rate (Gr) | Ingen restriktioner |
SAR | WB-SAR: 2,0 W/kg ved 15 minutters scanning Normal Operating Mode First Mode er ikke tilladt |
B1+rms | Ingen restriktioner |
Coil type | Ingen restriktioner |
Specielle forhold | Patienten kan scannes umiddelbart efter implantation |
Temperaturstigning | Maksimalt 2,3 grader ved Normal Operationg Mode i 15 min |
Artefaktdannelse | 28 - 36 mm fra implantatet |
Kilde | Produktinformation fra Edwards Lifesciences |
Gyldighed | Denne procedure kan anvendes som inspirationskilde til udvikling af egne sikkerhedsforskrifter |
Information | Procedure 074. Revision: 2.2 Tidligere udgaver af denne sikkerhedsvejledning er ikke gemt Udarbejdet af: Bo Haugaard Jørgensen Dato: 01.06.2022 Opdateres: Løbende, senest 01.01.2024 |