276_sikkerhed_stent

Stents fra MIT (M.I.Tech)

Fabrikat/model

Stents fra MIT (M.I.Tech)


Esophagus Stent

  • Covered  Esophagus Stent
  • Choostent
  • Esophagal cervical
  • Hanarostent
  • Hanarostent Valve
  • Hanarostent Benign
  • Gastro Seal
  • Hanarostent TTS Application
  • Hanarostent  Esophagus TTS


Biliary Stent

  • Hanarostent  Plumber
  • Hanarostent
  • NonCovered Biliary Stent
  • Large Cell
  • Benefit
  • Hybrid NonCovered Stent
  • Flap with Lasso


Trachea/Bronchium Stent

  • Hanarostent Trachea/Bronchium


Duodenal Stent

  • NonCovered Duodenum/Pylorus Stent
  • Single Layer
  • Dual Layer
  • Covered Duodenum/Pylorus Stent
  • Twynlay
  • Kim's Flare
  • Hanarostent


Colon Stent

  • Hanarostent

Statisk feltstyrke

1,5 Tesla og 3,0 Tesla

Scannerarkitektur

Ingen restrektion

Spatial gradient

Maksimalt 40 T/m alt. 4000 Gauss/cm (Kilde: IFU fra producenten)

Slew rate (Gr)

Ikke oplyst...

SAR

WB: 2,0 W/kg

Normal Operation Mode

B1+rms

Ingen restrektioner

Coil type

Ingen restrektioner

Artefakter

Billedartefakter må kunne forventes ved scanning i den region som involverer implantatet. Ved scanning på 3T-system med GE-sekvenser vil man således kunne se et susceptibilitetsartefakt/distortionsartefakt på op til 5 mm.

Kilde

IFU fra producenten

Gyldighed

Denne procedure kan anvendes som inspirationskilde til udvikling af egne sikkerhedsforskrifter

Information

Procedure 276. Revision: 1.5

Tidligere udgaver af denne sikkerhedsvejledning er ikke gemt

Udarbejdet af: Bo Haugaard Jørgensen

Dato: 12.08.2021

Opdateres: Løbende, senest 01.01.2023