wll_MatrixNEURO

MatrixNeuro

DePuy Synthes

Fabrikat/model

MatrixNeuro

DePuy Synthes

Bemærk

Producenten skriver...


It is recommended to thoroughly monitor patients undergoing MR scanning for
perceived temperature and/or pain sensations


Patients with impaired thermo regulation or temperature sensation should be
excluded from MR scanning procedures


Generally it is recommended to use a MRI system with low field strengths in the
presence of conductive implants. The employed specific absorption rate (SAR)
should be reduced as far as possible


Using the ventilation system may further contribute to reduce temperature increase in the body.

Statisk feltstyrke

Maksimalt 3,0 Tesla

Scannerarkitektur

Ingen restriktioner

Spatial gradient

Producenten skriver...

Non-clinical testing of worst case scenario in a 3 T MRI system did not reveal any
relevant torque or displacement of the construct for an experimentally measured
local spatial gradient of the magnetic field of 3.65 T/m


Kommentar...

Den maksimal spatial gradient må, qua ovenstående, forventes at ligge en del højere end 3,65 T/m. En lokal overvejelse må således lægges til grund for den endelige sikkerhedsvurdering...

Slew rate (Gr)

Ingen restriktioner

SAR

Maksimal WB-SAR: 2,0 W/kg pr. 15 min

Normal operating mode

B1+rms

Ingen restriktioner

Coil type

Ingen restriktioner

Temperaturstigning

Ved scanning med en WB-SAR på 2,0 W/kg...


1,5 Tesla: Maksimalt 11,3 grader efter 15 minutters scanning

3,0 Tesla: Maksimalt 8,5 grader efter 15 minutters scanning

Artefakter

Maksimalt 34 mm fra implantatet

Kilde

Sikkerhedsinformation fra producenten

Reference 1194_001, 1194_002, 1194_003 og 1194_004

Gyldighed

Denne procedure kan anvendes som inspirationskilde til udvikling af egne sikkerhedsforskrifter

Information

Procedure 1195. Revision: 1.5

Tidligere versioner af denne sikkerhedsvejledning er ikke gemt

Udarbejdet af: Bo Haugaard Jørgensen

Dato: 01.12.2023

Opdateres: Løbende, senest 01.01.2026